The tapered joints of modular hip implants are prone to fretting and crevice-corrosion. This can lead to total failure in under a year, especially for heavier, more active implant recipients. In this study, fractography of a failed Profemur Z implant showed that a life limiting fatigue crack was nucleated on the anterolateral surface of the implant's neck. The fatigue crack nucleation area appeared to have both more fretting damage and a higher corrosion rate than on other surfaces of the neck.
Details
- Fractography of a Neck Failure in a Double-Modular Hip Implant
- Williams, Jason (Author)
- Chawla, Nikhilesh (Author)
- Ira A. Fulton Schools of Engineering (Contributor)
-
Digital object identifier: 10.1016/j.csefa.2014.03.001
-
Identifier TypeInternational standard serial numberIdentifier Value2213-2902
Citation and reuse
Cite this item
This is a suggested citation. Consult the appropriate style guide for specific citation guidelines.
Williams, J. J., & Chawla, N. (2014). Fractography of a neck failure in a double-modular hip implant. Case Studies in Engineering Failure Analysis, 2(1), 45-50. doi:10.1016/j.csefa.2014.03.001